Menu-Driven User Set-up. IP65 (NEMA 4X) Enclosure. High/Low, and Alarm Relays. Dual Backlit LCD. Calibration Reminder Via Alarm Relay Setting. Isolated 0.4 to 20 mA Retransmission Output. RS485 Output. More
The CDCN-961 conductance, resistance or total dissolved solids (TDS) controller is designed for in-line monitoring of process solutions and water applications.
The CDCN-961 has a simple menu for easy set-up. Available relays can be configured for control, alarm or as a calibration reminder. The transmitter output can also be set from the keypad, eliminating the need for potentiometer or dip-switch settings. The calibration function automatically recognises the buffer solution resulting in a simple and straight forward system calibration.
SPECIFICATIONS:
Conductivity Range:
0.000 to 3.000, 30.00, 300.0 µS/cm
0.000 to 3.000, 30.00, 300.0 mS/cm
Resistivity Range:
0.3 to 30.0 MΩcm
3 to 300 MΩcm
TDS Range:
0.000 to 3.000, 30.00, 300.0 ppm
0.000 to 3.000, 30.00, 300.0 ppt
Resolution:
0.001/0.01/0.1 µS/cm
0.001/0.01/0.1 mS/cm
0.1 MΩcm
Accuracy: ±1% FS
Temperature Range: -0.5 to 130°C (31.1 to 266°F)
Temperature Resolution /Accuracy: 0.1°C/ ±0.5°C
Cell Constant: K = 0.01, 0.1, 1, 10
Line Resistance Compensation: 0.0 to 20.0 Ω
Temperature Sensor: Pt1000
Signal Output/Load: 0/4 to 20 mA/600Ω
Output Interface: RS485
Relay Specification: 3 SPST/250V 1A
Control Hysteresis Band: 0.1 to 10% FS
Frequency of Cleaning Time: 1 to 999 hours 1 to 999 seconds
EMC Standard: EN 50081-1/EN 50082-1
Operation Temperature: -10 to 50°C
Operation Humidity: 10 to 95%
Enclosure: NEMA 4X (IP65)
Dimensions: 100 W x 108 L x 136 Dia mm
Panel Cutout: 93.5 L x 93.5 W mm
Weight: 1650 g
Accessory:
Conductivity Electrode: K=1.0
Material: PPS/graphite
Test Range: 0 to 50.0 mS/cm
Temperature Range: 0 to 80°C
Temperature System: Pt1000
Cable: 5 metres

An Introduction to Industrial pH Measurement
What is pH? pH, which is the measurement of hydrogen ion concentration, describes the degree of acidity or alkalinity of a solution. pH measurement as well as pH control is a critical factor in a wide range of industrial applications – including pharmaceutical manufacturing, food and beverage production, dairy and agriculture, water treatment applications, and countless others.
pH control is often vital in ensuring the quality of a product...
* Fields marked with an asterisk are required
Accessories2

CDE-961

Conductivity Calibration Solution
Related products1