Portable Temperature and Humidity Data Loggers

Specify product selection
Variant does not exist

from € 125.10 excl. VAT
qty
OMEGA Engineering

Models for Temperature (OM-91) and Temperature/Humidity (OM-92). Free Software for Windows and Apple Mac. Immediate, Delayed (Specific Date/Time) or Pushbutton Start. Real-Time Mode. User-Configurable High and Low Alarms. LED Status Indicators. USB Interface for Fast Data Transfer. Low-Cost Calibration Options. Compact, Light Weight, Easy-to-Use. Long Battery Life. More

Item# (PN): OM-90_SERIES Shipping and Payment

The OM-90 Series are portable, battery operated, temperature (OM-91) and temperature/humidity (OM-92) data loggers. Logging is performed at a user configurable rate which is software selectable.

These data loggers are ideal for transportation applications where it is necessary to document that the material in-transit has stayed within particular environmental limits. High/low alarm limits can be set above or below which LED indicators will be activated so that the user can become aware that the desired transportation conditions have been exceeded. Each data logger can be assigned a 16 character name. This can be helpful in applications where multiple data loggers are used and there is a need to distinguish between them (for example based on location).

You may also use your data logger in Real Time Logging mode. This is a way to monitor temperature and humidity conditions directly from your PC or Mac instead of using the remote logging.

The data logger includes a very easy to use software application. The application is used for configuration of the logger, extraction of logger data, graphically viewing data, spread sheet analysis of data and saving data to file (typically in comma separated value CSV file format) for third party applications such as Microsoft Excel.

The graphical user interface is completely intuitive in terms of configuring new devices and extracting data. When the user has multiple devices requiring the same set-up then a pre-saved configuration file can be loaded into the application to help expedite the process and ensure all loggers get configured identically. The application does not require any driver installations to connect to the data logger. Just install the software, run it and then connect the data logger.

SPECIFICATIONS:

TEMPERATURE
Range: -30 to 80°C (-22 to 176°F)
Resolution: 0.01°C
Accuracy: ±0.3°C from 5 to 60°C (41 to 140°F)
Calibration Accuracy @ 25°C ±20°C: ±0.1°C

HUMIDITY (OM-92 Only)
Range: 0 to 95% RH
Resolution: 0.01% RH
Accuracy: ±3% over 20 to 80% RH range; ±5% RH below 20% RH or above 80% RH
Calibration Accuracy @ 25% RH: ±1.8% RH @ 25% RH

MEMORY
OM-91: 65,520 temperature measurements
OM-92: 65,520 temperature and 65,520 humidity measurements

Start Modes: Immediate start on disconnect from computer, delayed start (scheduled date/time) or pushbutton start
Recording Mode: Stop on memory full; when data logger operation is start/stopped by user intervention via button press, last memory location is retained
Recording Interval: 1 sec, 10 sec, 30 sec, 1 min, 10 min, 30 min, 1 hr (software selectable)
Software Compatibility: Windows 7 / 8 / 10 (32-bit and 64-bit), Apple Mac
Real Time Logging: Displays real time data when connected to a computer (PC, Apple Mac)
Real Time Clock: Time automatically synced to computer
Battery Backup: Continues to keep time when off
Accuracy: 0.50 sec/day (~3 min/year)
Alarms: High and Low alarms for temperature and humidity; dedicated alarm LED indicator
LED Indicators: Alarm and status LEDs
Device ID: Each data logger can have a 16 character name assigned
Connections: 90 cm USB Micro-B to computer cable (included)
Power: 3V lithium CR2450 battery included (user replaceable)
Battery Life: Up to 4 year battery life while logging
Bypass Mode: Unit is powered from USB when connected to conserve the battery
Dimensions: 53 H x 33 W x 15 mm D
Weight: 25 g